Gage R&R vs Type 1 gauge study

Short answer

A Type 1 study (VDA 5, Bosch Booklet 10) measures one reference part about 50 times to check the gauge’s own bias and repeatability, giving Cg and Cgk. A Gage R&R (AIAG MSA) measures about 10 parts by 2–3 operators 2–3 times each to find how much of the variation comes from the measurement system, giving %GRR and ndc. Do the Type 1 study first; if it fails, a Gage R&R makes no sense.

Side by side

AspectType 1 study (Cg, Cgk)Gage R&R (%GRR, ndc)
QuestionIs the gauge itself accurate and repeatable?How much of the observed variation is the measurement system?
PartsOne reference (master) part with known valueAbout 10 production parts covering the process range
OperatorsOneTwo or three
Measurements25–50 repeats10 parts × 3 operators × 2–3 trials
ResultsCg (repeatability), Cgk (with bias), bias t-test%GRR of study variation or tolerance, ndc, EV, AV, PV
Typical limitCg and Cgk ≥ 1.33%GRR < 10 % acceptable, 10–30 % conditional, > 30 % not acceptable; ndc ≥ 5
StandardVDA 5, Bosch Booklet 10AIAG MSA 4th edition

What each study can and cannot see

The Type 1 study isolates the instrument: one part, one person, ideal conditions. It finds bias and poor resolution but says nothing about operator influence or how the gauge behaves across real parts. Gage R&R adds operators and part-to-part variation, so it shows reproducibility and whether the gauge can tell good parts from each other, but it does not detect a constant bias, because every operator shares it.

Recommended order

Check resolution (at least 5 % of the tolerance), run the Type 1 study, then the Gage R&R. Add a linearity and bias study if the gauge is used over a wide range, and an attribute agreement analysis for visual or go/no-go checks.