Gauge calibration register online
Keep the list of gauges, test equipment and reference standards in the browser: ID, description, location, calibration interval, last calibration date and result. The register calculates the next due date and sorts the gauges by urgency.
Overdue gauges, gauges due within the warning period and gauges blocked after a failed calibration are flagged, so nothing is used outside its calibration status. The register prints as a landscape list for audits.
Worked example: Calibration Register
- 01Register7 gaugesinterval and last date
- 02Due datescalculatedlast + interval
- 03Status2 overdue, 1 blocked1 due in 6 days
- 04Actblock and planassess past results
01Method
Next due date = last calibration date + interval in months (end-of-month safe). A gauge is overdue after its due date, due soon within the warning period (30 days by default), blocked when the last result was not OK, and incomplete when the interval or date is missing.
A status date can be set to check the register as of an audit date; by default it is today.
02When to use it
Use it in small plants and labs without a calibration software, to prepare an ISO 9001 or IATF 16949 audit, and to plan the calibration work of the coming weeks.
PDF report in one click
Export a landscape PDF with the register sorted by urgency and the status of each gauge.
Private: runs in your browser
Your register is stored in your browser only.
How to keep the register
- Give every gauge an ID and a label with the next due date.
- Set the interval from the manufacturer, the use and the history.
- Enter each calibration with its date and result.
- Check the register weekly for gauges due within 30 days and plan them.
- Block failed and overdue gauges and assess their past results.
Read the guide to gauge calibration management, prove fitness for the tolerance with a Gage R&R and an uncertainty budget. All tools: quality inspection.
Frequently asked questions
›What does ISO 9001 require for calibration?
Clause 7.1.5.2 requires measuring equipment used to give evidence of conformity to be calibrated or verified at specified intervals against standards traceable to international or national standards, identified so its status is visible, and protected. When it is found unfit, the organisation must assess the validity of previous results.
›How long should the calibration interval be?
Start with the manufacturer’s recommendation or 12 months, then adjust from the history: extend it after several calibrations without adjustment, shorten it after a failure or for heavily used gauges.
›What do I do when a gauge fails calibration?
Block it, assess which parts were measured with it since the last good calibration, decide whether they are affected and record the decision. IATF 16949 7.1.5.2.1 also asks to inform the customer if suspect product was shipped.
›Is a calibration register enough for IATF 16949?
IATF 16949 7.1.5.2.1 also asks for a record of every gauge including employee-owned ones, calibration results, as-found condition, and actions for out-of-tolerance findings. Use the register as the overview and keep the certificates with it.