Measurement uncertainty calculator after the GUM
Set up an uncertainty budget for a measurement: list the contributors such as the calibration of the reference, repeatability, resolution, temperature, operator and form deviation, choose a distribution for each, and get the combined and expanded uncertainty.
With the tolerance entered, the tool shows the ratio 2U / T, judges whether the measurement process is suitable and calculates guard-banded acceptance limits. A chart shows which contributor to reduce first.
Worked example: Measurement Uncertainty Calculator
- 01Contributors6ring, repeatability, resolution …
- 02Combineu = 0.9 µm√Σu²
- 03ExpandU = 1.8 µmk = 2
- 04Judge2U / T = 17 %suitable, ≤ 30 %
01Method
Each value is converted to a standard uncertainty: normal values are divided by their coverage factor (1 for a standard deviation, 2 for most certificates), rectangular half-widths by √3, triangular by √6 and U-shaped by √2. They are combined as the root sum of squares and multiplied by the coverage factor k.
Sensitivity coefficients are taken as 1, so all values must be in the unit of the measurand. Guard-banded limits are the specification limits moved inwards by U.
02When to use it
Use it when a measurement result is close to a tolerance limit, to decide whether a gauge is suitable for a tight tolerance, to set guard bands for final inspection, and for accredited laboratories and VDA 5 studies.
PDF report in one click
Export a PDF with the measurement task, the full budget, the chart and the method.
Private: runs in your browser
The budget stays in your browser.
How to set up an uncertainty budget
- Define the measurand and the measurement method exactly.
- List every contributor: reference standard, repeatability, resolution, temperature, operator, form of the part, fixture.
- Give each a value and a distribution: certificate values with their k, limits as half-widths.
- Read the combined and expanded uncertainty and the largest contributors.
- Compare 2U with the tolerance and use guard bands where U is significant.
Worked example
An air gauge measures a bore Ø22 H7 (tolerance 21 µm). Six contributors give a combined standard uncertainty of 0.9 µm and an expanded uncertainty U = 1.8 µm at k = 2. 2U / T = 17 %, so the measurement process is suitable; the largest contributor is the form deviation of the bore. See also measurement uncertainty in the glossary, the Gage R&R and the type 1 study.
Frequently asked questions
›What is the difference between measurement uncertainty and Gage R&R?
A Gage R&R measures repeatability and operator influence on real parts. An uncertainty budget adds every other source: calibration of the standard, resolution, temperature, form of the part, and so on. VDA 5 and ISO 22514-7 combine both into one measurement process capability.
›Which distribution do I choose?
Normal for values from a certificate (divide by its k) or a standard deviation (k = 1). Rectangular for limits without more information, such as resolution or a specified maximum error. Triangular when values near the centre are more likely, U-shaped for cyclic effects such as temperature swings.
›What is a guard band?
A narrower acceptance zone inside the tolerance, reduced by the expanded uncertainty at each side. A part measured inside the guard-banded limits is in tolerance with the stated confidence.
›What is a good ratio of uncertainty to tolerance?
2U / T below 30 % is a common limit for a suitable measurement process (VDA 5 style); some customers ask for 20 % or less for special characteristics.