Measurement uncertainty
range in which the true value lies
U = k · √(Σ uᵢ²)
Measurement uncertainty describes the range around a measured value within which the true value lies with a stated probability. It is estimated by listing all influences (gauge, reference, temperature, operator, part, repeatability), giving each a standard uncertainty uᵢ, combining them as the root sum of squares and multiplying by a coverage factor k, usually 2 for about 95 %.
The GUM (ISO/IEC Guide 98-3) defines the method; VDA Volume 5 and ISO 22514-7 apply it to measurement process capability in production. Near the tolerance limits, uncertainty decides whether a part can be called conforming.
Calculate it
Gage R&R (MSA): Crossed study with ANOVA: repeatability, reproducibility, %GRR, % tolerance and ndc.
Related terms
- MSA: measurement system analysis
- Calibration: comparing an instrument with a reference
- %GRR: gauge repeatability and reproducibility